SabreAlert3
Category: Characterization > Monitors > Radiation Monitors
Reference # : Model No :
Lightweight, battery powered, alpha air monitor that can be used as a portable workplace monitor or a portable CAM for emergency response assessments.
Benefits
Weighs 5.9 lbs.
10-Hour battery life.
Uses alpha peak-shape-fitting for individual measurements of nuclide counts–significantly reducing false alarms.
Performs both fast-responding (Acute) and high sensitivity (Chronic) measurements.
Can be plumbed to house vacuum or external pump for higher flow rates and improved sensitivity.
Acute and Chronic dose, concentration, and flow logging as well as spectrum logging.
Wireless 802.11b RadNet Output Option for remote monitoring.
Typical sensitivity of 1.0 DAC-h (Chronic) and 30 DAC-h (Acute) with less than one false alarm per 2080 hours of operation.
Radon Mode Option displays PAEC and also indicates radon progeny equilibrium.
American or SI units.
Limitations
Detector:
Solid-state ion-implanted silicon (450 mm² active area)
Pump:
diaphragm-type, 7.0 LPM (typical)
Filter:
37 mm Speclon™ 1.5 µm PTFE membrane or compatible 37mm PTFEfilter (25 mm collection area)
MCA:
1024-channel ADC binned to 256 channel spectrum
Processor:
Windows CE-based PDA, 533 MHZ Intel X-Scale processor
Max Count Rate:
600,000 cpm
Calibration:
Electro-plated stainless steel source required for effciency calibration, 37mm diameter with 25mm active area.
Energy Range:
1.0 - 8.0 MeV (1.0 - 10 MeV in Radon Mode)
Battery powered:
8.4 V Li-Ion, 6.3 Ah 10-hour run-time, 4-hour charge time
Weight:
5.9 lbs (2.7 kg)
Dimensions:
9”w x 9”h x 4”d (24 x 24 x 10cm)
Temperature:
0 to 122 °F (-20 to 50°C)
Humidity:
5 to 100% (non condensing)